Remote Control Commands
R&S
®
SMA100B
424User Manual 1178.3834.02 ─ 05
Return values:
<CalInfoText> string
Example:
CAL:ALL:INF?
"Instrument is calibrated, no adjustment required."
"UNCAL, instrument is warming up."
"UNCAL, Please perform full adjustment after warming up."
"UNCAL, Please perform full adjustment."
Usage: Query only
Manual operation: See "Information" on page 696
:CALibration<hw>:ALL:TEMP?
Queries the temperature deviation compared to the calibration temperature.
Suffix:
<hw>
.
[1]
Optional suffix
Return values:
<Temperature> string
Example:
CALibration:ALL:TEMP?
// "+39.00 K"
Usage: Query only
Manual operation: See "Temperature Offset Since Last Full Adjustment"
on page 696
:CALibration<hw>:ALL:TIME?
Queries the time elapsed since the last full adjustment.
Return values:
<Time> string
Example:
CAL:ALL:TIME?
// "22 days"
Usage: Query only
Manual operation: See "Time Since Last Full Adjustment" on page 696
:CALibration:DATA:FACTory:DATE?
Queries the date of the last factory calibration.
Return values:
<Date> string
Example:
CAL:DATA:FACT:DATE?
// "2016-01-01"
CALibration Subsystem