Remote Control Commands
R&S
®
SMA100B
471User Manual 1178.3834.02 ─ 05
Example:
SENS:SWE:HCOP:DEV:LANG:CSV:ORI VERT
selects horizontal orientation
SENS:SWE:HCOP:DATA?
queries the measurement data of the current traces
Response:
#2681009500000;-9.5; 1019000000;-9.7; 1028500000;-6.3;
1038000000;-2.5;
for vertical representation the length of the data block is 68
the first power value, the second row contains the second fre-
quency value of the active trace followed by the second power
value, and so on. The rows end with a new line (each counts as
one character).
Note: if more than one trace is active, the first row also contains
the value pairs of the second active trace, and so on.
Usage:
Query only
:SENSe[:POWer]:SWEep:HCOPy:DEVice <Device>
Defines the output device. The setting is fixed to FILE, i.e. the hardcopy is stored in a
file.
Parameters:
<Device> FILE | PRINter
*RST: FILE
Example:
SENS:SWE:HCOP:DEV FIL
Selects output device file.
:SENSe[:POWer]:SWEep:HCOPy:DEVice:LANGuage <Language>
Selects the bitmap graphic format for the screenshot of the power analysis trace.
In addition, ASCII file format *.csv is offered. If file format *.csv is selected, the
trace data is saved as an ASCII file with comma separated values. It is also possible to
directly retrieve the data using command :SENSe[:POWer]:SWEep:HCOPy:DATA?
on page 470
Parameters:
<Language> BMP | JPG | XPM | PNG | CSV
*RST: BMP
Example:
SENS:SWE:HCOP:DEV:LANG BMP
Selects output format *.bmp.
Manual operation: See " Format - Power Analysis " on page 255
:SENSe[:POWer]:SWEep:HCOPy:DEVice:LANGuage:CSV:DPOint <DPoint>
Defines which character is used as the decimal point of the values, either dot or
comma.
Power Sensor Measurement Subsystems