Remote Control Commands
R&S
®
SMA100B
531User Manual 1178.3834.02 ─ 05
Parameters:
<Mode> AUTO | EXTernal | EGATe | SINGle | ESINgle
*RST: AUTO
Example: See Example "Perform pulse modulation" on page 529.
Manual operation: See "Trigger Mode" on page 86
[:SOURce]:PULM[:INTernal][:TRAin]:TRIGger:IMMediate
If [:SOURce<hw>]:PULM:TRIGger:MODE SINGle, triggers the pulse generator.
Example: See Example "Perform pulse modulation" on page 529.
Manual operation: See "Execute Single Trigger" on page 89
[:SOURce<hw>]:PULM:PERiod <Period>
Sets the period of the generated pulse, that means the repetition frequency of the inter-
nally generated modulation signal.
Parameters:
<Period> float
The minimum value depends on the installed options
R&S SMAB-K22 or R&S SMAB-K23
Range: 20E-9 to 100
Increment: 5E-9
*RST: 10E-6
Example: See Example "Perform pulse modulation" on page 529.
Manual operation: See "Pulse Period" on page 88
[:SOURce<hw>]:PULM:WIDTh <Width>
Sets the width of the generated pulse, that means the pulse length. It must be at least
20ns less than the set pulse period.
Parameters:
<Width> float
Range: 20E-9 to 100
Increment: 10E-9
*RST: 2E-6
Example: See Example "Perform pulse modulation" on page 529.
Manual operation: See "Pulse Width" on page 88
[:SOURce<hw>]:PULM:DELay <Delay>
Sets the pulse delay.
SOURce Subsystem