Remote Control Commands
R&S
®
SMA100B
619User Manual 1178.3834.02 ─ 05
Default unit: dBm
Example: See Example "Create an RF list and activate the list mode"
on page 612.
Manual operation: See "Edit List Mode Data" on page 190
[:SOURce<hw>]:LIST:POWer:POINts?
Queries the number (points) of level entries in the selected list.
Return values:
<Points> integer
Range: 0 to INT_MAX
*RST: 0
Example: See Example "Create an RF list and activate the list mode"
on page 612.
Usage: Query only
[:SOURce<hw>]:LIST:TRIGger:EXECute
Starts the processing of a list in list mode.
Example:
See Example "Create an RF list and activate the list mode"
on page 612.
Usage: Event
Manual operation: See " Execute Single" on page 185
[:SOURce<hw>]:LIST:TRIGger:SOURce <Source>
Selects the trigger source for processing lists.
The names of the parameters correspond to those in sweep mode. SCPI standard
uses other names for the parameters; these names are also accepted by the instru-
ment. The SCPI names should be used if compatibility is an important consideration.
For an overview, see the following table:
Rohde & Schwarz name SCPI name Command in manual control
AUTO IMMediate
MODE AUTO
SINGle BUS MODE SINGLE or STEP
EXTernal EXTernal MODE EXT TRIG SINGLE or
EXT TRIG STEP
Parameters:
<Source> SINGle | BUS | AUTO | EXTernal | IMMediate
SOURce Subsystem