Remote Control Commands
R&S
®
SMA100B
651User Manual 1178.3834.02 ─ 05
Parameters:
<Logarithmic> float
The unit is mandatory.
Range: 0.01 to 100
Increment: 1E-3
*RST: 1
Default unit: PCT
Example: See Example "Setup an RF frequency or power sweep"
on page 643.
Manual operation: See " Step Linear/Step Logarithmic " on page 177
[:SOURce<hw>]:SWEep[:FREQuency]:STEP[:LINear] <Linear>
Sets the step width for linear sweeps.
See Chapter 6.2.1, "Correlating Parameters in Sweep Mode", on page 164.
Omit the optional keywords so that the command is SCPI-compliant.
Parameters:
<Linear> float
Range: 0.001 Hz to (STOP - STARt)
Increment: 0.01
Example: See Example "Setup an RF frequency or power sweep"
on page 643.
Manual operation: See " Step Linear/Step Logarithmic " on page 177
[:SOURce<hw>]:SWEep:RESet[:ALL]
Resets all active sweeps to the starting point.
Usage:
Event
Manual operation: See " Reset Sweep " on page 175
[:SOURce<hw>]:SWEep[:FREQuency]:TIME <Time>
Sets the duration of a frequency ramp sweep step.
Parameters:
<Time> float
Range: 0.01 to 100
Increment: 1E-4
*RST: 0.015
Default unit: s
Example:
SOURce1:SWEep:FREQuency:SPACing RAMP
SOURce1:SWEep:FREQuency:TIME 0.020
SOURce Subsystem