Remote Control Commands
R&S
®
SMA100B
661User Manual 1178.3834.02 ─ 05
:SYSTem:SAV <Pathname>
Saves the current instrument settings to a file with defined filename.
Setting parameters:
<Pathname> string
Example:
SYSTem:SAV
"/var/user/Instrument_settings.svrcltxt"
Usage: Setting only
:SYSTem:RCL <Pathname>
Loads a file with previously saved instrument settings.
Setting parameters:
<Pathname> string
Example:
SYSTem:RCL
"/var/user/Instrument_settings.svrcltxt"
Usage: Setting only
:SYSTem:PROTect<ch>[:STATe] <State>[, <Key>]
Activates and deactivates the specified protection level.
Suffix:
<ch>
.
Indicates the protection level.
See also "Protection" on page 332
Parameters:
<State> 0 | 1 | OFF | ON
*RST: n.a. (factory preset: 1)
Setting parameters:
<Key> integer
The respective functions are disabled when the protection level
is activated. No password is required for activation of a level.
A password must be entered to deactivate the protection level.
The default password for the first level is 123456.
This protection level is required to unlock internal adjustments
for example.
Example: To activate protection level:
SYSTem:PROTect1:STATe 1
Internal adjustments or hostname cannot be changed.
To unlock protection level 1:
SYSTem:PROTect1:STATe 0,123456
Internal adjustments are accessible.
Manual operation: See "Protection Level/Password" on page 333
SYSTem Subsystem