Remote Control Commands
R&S
®
SMA100B
684User Manual 1178.3834.02 ─ 05
EAUTo
An external signal triggers the sweep. When one sweep is fin-
ished, the next sweep starts.
A second trigger event stops the sweep at the current frequency,
a third trigger event starts the trigger at the start frequency, and
so on.
*RST: AUTO
Example: See Example "Setup an LF sweep" on page 599
Usage: Setting only
Manual operation: See "Mode" on page 172
:TRIGger<hw>:FSWeep[:IMMediate]
:TRIGger<hw>:PSWeep[:IMMediate]
:TRIGger<hw>:LFFSweep:IMMediate
:TRIGger<hw>[:SWEep][:IMMediate]
Performs a single sweep and immediately starts the activated, corresponding sweep:
●
FSWeep - RF frequency
●
PSWeep - RF level
●
LFFSweep - LF frequency
●
SWEep - all sweeps
Effective in the following configuration:
●
TRIG:FSW|LFFS|PSW|[:SWE]:SOUR SING
●
SOUR:SWE:FREQ|POW:MODE AUTO or SOUR:LFO:SWE:[FREQ:]MODE AUTO
Alternatively, you can use the IMMediate command instead of the respective SWEep:
[FREQ:]|POW:EXECute command.
Example:
TRIG
Starts all active sweeps.
Usage: Event
Manual operation: See " Execute Single Sweep " on page 175
:TRIGger<hw>:LFFSweep
Executes an LF frequency sweep in the following configuration:
●
TRIG:LFFS:SOUR SING
●
LFO:SWE:MODE AUTO
Example:
LFO:SWE:MODE AUTO
TRIG:LFFS:SOUR SING
TRIG:LFFSweep
Usage: Event
TRIGger Subsystem