Maintenance
R&S
®
SMA100B
696User Manual 1178.3834.02 ─ 05
If the termination resistor is missing, a second warning message appears.
Remote command:
:CALibration:ALL[:MEASure]? on page 423
Last Full Adjustment
Displays the date of the last fully performed adjustment.
Remote command:
:CALibration<hw>:ALL:DATE? on page 423
Time Since Last Full Adjustment
Displays the elapsed time since the last full adjustment.
Remote command:
:CALibration<hw>:ALL:TIME? on page 424
Temperature Offset Since Last Full Adjustment
Displays the temperature difference, comparing the temperature at the beginning of the
last adjustment to the temperature at the beginning of the present adjustment.
If the temperature deviates more than ± 5 K, the background of the parameter turns
yellow.
Remote command:
:CALibration<hw>:ALL:TEMP? on page 424
Information
Displays information to the current adjustment state.
Remote command:
:CALibration<hw>:ALL:INFormation? on page 423
Performing Maintenance Tasks