Remote Control Commands
R&S
®
SMB100A
396Operating Manual 1407.0806.32 ─ 13
Parameters:
<LTerminator> STANdard | EOI
EOI
The terminator must be sent together with the line message EOI
(End of Line). This setting is recommended for binary block trans-
missions where a character could coincidentally have the value LF
(Line Feed) but is not intended as the terminator. This setting must
be selected for block data with undefined length.
STANdard
An LF (Line Feed) is recognized as the terminator regardless of
whether it is sent with or without EOI.
*RST: STANdard
Example:
SYST:COMM:GPIB:LTER EOI
only a character which is sent simultaneously with the line mes-
sage EOI is accepted as the terminator.
:SYSTem:COMMunicate:GPIB:RESource?
The command queries the visa resource string for remote control via the GPIB interface.
This string is used for remote control of the instrument.
To change the GPIB address, use the command SYST:COMM:GPIB:ADDR.
Return values:
<Resource> string
Example:
SYST:COMM:GPIB:RES
queries the VISA resource string.
Response: GPIB::28::INSTR
Usage: Query only
Manual operation: See "Visa Resource Strings" on page 96
:SYSTem:COMMunicate:GPIB[:SELF]:ADDRess <Address>
Sets the GPIB address.
Parameters:
<Address> integer
Range: 1 to 30
*RST: 28
Example:
SYST:COMM:GPIB:ADDR 14
sets GPIB address.
Manual operation: See "GPIB channel address" on page 95