Loading...
Amplitude Modulation
Characteristics Included in Specified min.
value
Measured
value
Specified
max. value
Unit Uncert.
AM Setting Uncertainty
Intern
Page 1.31
Intern, 5%
3.80 6.20 % 0.10
Intern, 10%
8.60 11.40 % 0.10
Intern, 20%
18.20 21.80 % 0.10
Intern, 40%
37.40 42.60 % 0.10
Intern, 60%
56.60 63.40 % 0.10
Intern, 80%
75.80 84.20 % 0.10
Extern
Extern, 5%
3.80 6.20 % 0.10
Extern, 10%
8.60 11.40 % 0.10
Extern, 20%
18.20 21.80 % 0.10
Extern, 40%
37.40 42.60 % 0.10
Extern, 60%
56.60 63.40 % 0.10
Extern, 80%
75.80 84.20 % 0.10
AM 80 %
80%, 100 kHz
78.20 81.80 % 0.10
80%, 23.4375 MHz
78.20 81.80 % 0.10
80%, 23.438 MHz
75.80 84.20 % 0.10
80%, 374.9 MHz
75.80 84.20 % 0.10
80%, 375 MHz
75.80 84.20 % 0.10
80%, 529.9 MHz
75.80 84.20 % 0.10
80%, 530 MHz
75.80 84.20 % 0.10
80%, 1449.9 MHz
75.80 84.20 % 0.10
80%, 1450 MHz
75.80 84.20 % 0.10
80%, 2120.9 MHz
75.80 84.20 % 0.10
80%, 2121 MHz
75.80 84.20 % 0.10
80%, 2999.9 MHz
75.80 84.20 % 0.10
80%, 3000 MHz
75.80 84.20 % 0.10
80%, 3999.9 MHz
75.80 84.20 % 0.10
80%, 4000 MHz
75.80 84.20 % 0.10
80%, 6000 MHz
75.80 84.20 % 0.10
Model
SMB100A
Serial no.
100063
Item
Signal Generator
Order no.
1406.6000k03
Cal. date
2015-04-16
Page
35 of 50