Instrument Function
R&S
®
SMBV100A
285Operating Manual 1176.8016.02 ─ 17
"Extern Single"
Sets a single sweep cycle. The sweep is triggered by an external trig-
ger signal.
Refer to the description of the rear panel for information about the
connectors for external trigger signal input (see Chapter 2.2, "Rear
Panel Tour", on page 28).
Example:
SOUR:SWE:POW:MODE AUTO
TRIG:PSW:SOUR EXT
SOUR:POW:MODE SWE (External trigger)
"Extern Step"
Sets a step-by-step sweep cycle. Each sweep step is triggered by an
external trigger signal (trigger source as described under "Extern Sin-
gle"). The step width corresponds to the step width of the rotary knob.
Example:
SOUR:SWE:POW:MODE STEP
SOUR:SWE:POW:STEP 0.5
TRIG:PSW:SOUR EXT
SOUR:POW:MODE SWE (External trigger)
"Extern Start/Stop"
Sets an automatically repeated sweep cycle that is started, stopped
and restartet by subsequent external trigger events.
The first external trigger signal starts the sweep (Start).
The next external trigger signal stops the sweep at the current fre-
quency (Stop).
The third external trigger signal starts the sweep at the start fre-
quency (Start).
Refer to the description of the rear panel for information about the
connectors for external trigger signal input (see Chapter 2.2, "Rear
Panel Tour", on page 28).
Example:
SOUR:SWE:POW:MODE AUTO
TRIG:PSW:SOUR EAUT
SOUR:POW:MODE SWE (External trigger)
Remote command:
[:SOURce<hw>]:SWEep:POWer:MODE on page 814
:TRIGger<hw>:PSWeep:SOURce on page 846.
[:SOURce<hw>]:POWer:MODE on page 791
Reset Sweep - Level Sweep
Resets the sweep. The start level is set and the next sweep starts from there.
Remote command:
[:SOURce<hw>]:SWEep:RESet[:ALL] on page 817