Instrument Function
R&S
®
SMBV100A
286Operating Manual 1176.8016.02 ─ 17
Execute Single Sweep - Level Sweep
Triggers the sweep manually. A manual sweep can only be triggered if "Mode Single"
is selected.
Example:
SOUR:SWE:POW:MODE AUTO
TRIG:PSW:SOUR SING
SOUR:POW:MODE SWE
SOUR:SWE:EXEC
Remote command:
[:SOURce<hw>]:SWEep:POWer:EXECute on page 813
:TRIGger<hw>:PSWeep[:IMMediate] on page 847
:TRIGger<hw>[:SWEep][:IMMediate] on page 848
Use Digital Attenuation
Activates the use of digital attenuation during RF power sweep.
Remote command:
[:SOURce<hw>]:SWEep:POWer:DATTenuator:STATe on page 813
Start Level - Level Sweep
Sets the start level.
Remote command:
[:SOURce<hw>]:POWer:STARt on page 794
Stop Level - Level Sweep
Sets the stop level.
Remote command:
[:SOURce<hw>]:POWer:STOP on page 796
Current Level - Level Sweep
Displays the current level.
If "Step" is set, the level for the next level step of the sweep is entered here.
Remote command:
[:SOURce<hw>]:POWer:MANual on page 790
Shape - RF Level Sweep
Selects the cycle mode for a sweep sequence (shape).
"Sawtooth"
One sweep runs from the start level to the stop level. The subsequent
sweep starts at the start level again, i.e. the shape of sweep
sequence resembles a sawtooth.
"Triangle"
One sweep runs from start to stop level and back, i.e. the shape of
the sweep resembles a triangle. Each subsequent sweep starts at the
start level again.
Remote command:
[:SOURce<hw>]:SWEep:POWer:SHAPe on page 815
RF Block

Table of Contents