Remote Control Commands
R&S
®
SMBV100A
629Operating Manual 1176.8016.02 ─ 17
Parameters:
<Step> float
Range: -80 to 80
Increment: 0.01
*RST: 0
Default unit: dB
Example:
BB:ARB:MCAR:EDIT:CARR:POW -80dB
sets a power of -80 dB for the carriers in the carrier range.
BB:ARB:MCAR:EDIT:CARR:POW:STEP 1 dB
the power is incremented by 1dB for each carrier, i.e. the first
carrier has -80dB, the second -79dB, etc.
Manual operation: See " Gain Step " on page 433
[:SOURce<hw>]:BB:ARBitrary:MCARrier:EDIT:CARRier:POWer[:STARt] <Start>
The command sets the power for the individual carriers in the defined carrier range. If
the command :BB:ARB:MCAR:EDIT:CARR:POW:STEP is used to define a step width,
the power entered here applies only to the starting carrier. The power of the remaining
carriers is stepped up or down by the power specified in
the :BB:ARB:MCAR:EDIT:CARR:POW:STEP command.
Parameters:
<Start> float
Range: -80 to 0
Increment: 0.01
*RST: 0
Default unit: dB
Example:
BB:ARB:MCAR:EDIT:CARR:POW -50 dB
sets the power of the carriers in the carrier range to -50 dB.
Manual operation: See " Gain Start " on page 433
[:SOURce<hw>]:BB:ARBitrary:MCARrier:EDIT:CARRier:STARt <Start>
[:SOURce<hw>]:BB:ARBitrary:MCARrier:EDIT:CARRier:STOP <Stop>
The command selects the last carrier in the carrier range to which the settings with
the :BB:ARB:MCAR:EDIT:CARR:.. commands shall apply.
Parameters:
<Stop> integer
Range: 0 to 511
*RST: 0
Example:
BB:ARB:MCAR:EDIT:CARR:STOP 4
the carrier range stops at carrier 4.
Manual operation: See " Carrier Stop " on page 433