Remote Control Commands
R&S
®
SMBV100A
632Operating Manual 1176.8016.02 ─ 17
Manual operation: See "Save/Recall Frame" on page 425
[:SOURce<hw>]:BB:ARBitrary:MCARrier:SETTing:STORe:FAST <Fast>
Determines whether the instrument performs an absolute or a differential storing of the
settings.
Enable this function to accelerate the saving process by saving only the settings with
values different to the default ones.
Note: This function is not affected by the "Preset" function.
Parameters:
<Fast> 0 | 1 | OFF | ON
*RST: 1
Manual operation: See "Fast Save" on page 162
[:SOURce<hw>]:BB:ARBitrary:MCARrier:TIME <Time>
Sets the user-defined signal period. This setting is only possible for Signal Period
Mode User (BB:ARB:MCAR:TIME:MODE USER).
Parameters:
<Time> float
Range: 0 to 1E9
Increment: 1E-9
*RST: 0
Default unit: s
Example:
BB:ARB:MCAR:TIME:MODE USER
selects Signal Period Mode User.
BB:ARB:MCAR:TIME 10 s
sets a signal period of 10 seconds
Manual operation: See "Signal Period" on page 428
[:SOURce<hw>]:BB:ARBitrary:MCARrier:TIME:MODE <Mode>
Selects the mode for calculating the resulting signal period of the multi carrier wave-
form. The resulting period is always calculated for all carriers in the carrier table irre-
spective of their state (ON/OFF).
SOURce Subsystem