Remote Control Commands
R&S
®
SMBV100A
673Operating Manual 1176.8016.02 ─ 17
Example:
BB:DM:PRAM:SHAP COS
selects a cosine-shaped rise and fall for the transmitted power
edge.
Manual operation: See "Ramp Function" on page 370
[:SOURce<hw>]:BB:DM:PRAMp:SOURce <Source>
Sets the source for the power ramp control signals.
Parameters:
<Source> INTernal
INTernal
The internal control signals LEV_ATT and BURST are used.
Control signals are stored in dedicated lists. When remote con-
trol is in use, the list to be used is defined, selected and enabled
with the aid of SOURce:BB:DM:CLISt:... commands.
*RST: INTernal
Example:
BB:DM:PRAM:SOUR INT
enables the use of internal control signals for power ramp con-
trol.
[:SOURce<hw>]:BB:DM:PRAMp:TIME <Time>
Sets the power ramping rise time and fall time for a burst.
Parameters:
<Time> float
Range: 0.25 to 16
Increment: 0.01
*RST: 1
Default unit: symbol
Example:
BB:DM:PRAM:TIME 2
sets a time of 2 symbols for the edges to rise and fall.
Manual operation: See "Ramp Time" on page 370
[:SOURce<hw>]:BB:DM:PRAMp[:STATe] <State>
The command enables or disables power ramping.
Parameters:
<State> 0 | 1 | OFF | ON
*RST: 0
Example:
BB:DM:PRAM:STAT ON
switches power ramping on.
Manual operation: See "State" on page 370
SOURce Subsystem