Remote Control Commands
R&S
®
SMBV100A
679Operating Manual 1176.8016.02 ─ 17
[:SOURce<hw>]:BB:DM:DLISt:DATA <Data>
[:SOURce<hw>]:BB:DM:DLISt:DATA? [<Start>[, <Count>]]
The Setting command sends the bit data to the data list selected with the com-
mand :BB:DM:DLISt:SELect. Any existing content in the data list is overwritten.
This command only writes data into the data section of the file.
This command sends the bit data to the selected data list, which is overwritten.
The data can also be sent as block data in binary or PACKed format, each byte being
interpreted as 8 data bits. Use the command :SYSTem:COMMunicate:GPIB:
LTERminator EOI to set the termination character mode to ‘EOI control data mes-
sage only’. The command causes that a random LF in the data sequence is not inter-
preted as end, i.e. prematurely terminating the data transmission. The com-
mand ...LTER STAN resets the mode.
The query reads out the data part of the data list. If the query is expanded by using the
two parameters <start> and <length>, the list is read out in smaller sections. Start and
Length are expressed in bits. Without the parameters, the total length is always read
out starting from address 1. The command :FORMat ASCii | PACKed can be used
to select the data format. The byte sequence is defined in the IEC bus standard (read/
write most significant byte first). *RST has no effect on data lists.
Parameters:
<Data> integer
Query parameters:
<Start> integer
Range: 1 to 2147483647
<Count> integer
Range: 1 to 2147483647
Example:
BB:DM:DLIS:SEL 'dlist1'
Selects data list dlist1. If the file does not yet exist, it is created.
BB:DM:DLIS:DATA 1,1,1,0,0,0,1,1,0,1...
Sends the specified data to file dlist1. Any data already present
is overwritten.
Example:
BB:DM:DLIS:SEL 'dlist1'
Selects data list dlist1.
FORM ASCI
Selects ASCII data transmission format.
BB:DM:DLIS:DATA? 2048,1024
Queries the data starting at bit 2048 for 1024 bits.
Manual operation: See "Data List Editor" on page 364