Remote Control Commands
R&S
®
SMBV100A
710Operating Manual 1176.8016.02 ─ 17
Return values:
<Maximum> float
Range: 0 to 16777215
Increment: 1E-3
Example:
BB:MCCW:TRIG:OUTP:DEL:FIX ON
restricts the marker signal delay setting range to the dynamic
range.
BB:MCCW:TRIG:OUTP:DEL:MAX
queries the maximum of the dynamic range.
Response: 2000
the maximum for the marker delay setting is 2000 samples.
Usage: Query only
Manual operation: See "Current Range without Recalculation" on page 448
[:SOURce<hw>]:BB:MCCW:TRIGger:OUTPut<ch>:DELay:MINimum?
Queries the minimum marker delay for setting :BB:MCCW:TRIGger:OUTPut:
DELay:FIXed ON.
Return values:
<Minimum> float
Range: 0 to 16777215
Increment: 1E-3
*RST: 0
Example:
BB:MCCW:TRIG:OUTP:DEL:FIX ON
restricts the marker signal delay setting range to the dynamic
range.
BB:MCCW:TRIG:OUTP:DEL:MIN
queries the minimum of the dynamic range.
Response:0
the minimum for the marker delay setting is 0 samples.
Usage: Query only
Manual operation: See "Current Range without Recalculation" on page 448
[:SOURce<hw>]:BB:MCCW:TRIGger:OUTPut<ch>:MODE <Mode>
Defines the signal for the selected marker output.
SOURce Subsystem