Remote Control Commands
R&S
®
SMBV100A
730Operating Manual 1176.8016.02 ─ 17
Example:
CORR:CSET '/var/user/ucor1'
selects the table ucor1.
CORR ON
activates level correction. Correction is performed using the
table ucor1.
Manual operation: See " User Cor. Data - User Correction " on page 233
[:SOURce<hw>]:CORRection[:STATe] <State>
Activates/deactivates level correction. Level correction is performed using the table
which has been selected with the command [:SOURce<hw>]:CORRection:CSET[:
SELect].
Parameters:
<State> 0 | 1 | OFF | ON
*RST: 0
Example:
SOUR:CORR:CSET '/var/user/ucor1'
selects the table ucor1.
SOUR:CORR ON
activates user correction.
Manual operation: See " State - User Correction " on page 233
[:SOURce<hw>]:CORRection:VALue?
Queries the current value for user correction.
Return values:
<Value> float
Range: -100 to 100
Increment: 0.01
*RST: 0
Example:
CORR:VAL?
queries the value currently used for level correction.
Response: -3
the correction value is - 3 dB.
Usage: Query only
Manual operation: See " User Correction Value - User Correction " on page 233
[:SOURce<hw>]:CORRection:ZERoing:STATe <State>
Activates the zeroing procedure before filling the user correction data acquired by a
sensor.
Parameters:
<State> 0 | 1 | OFF | ON
*RST: 1