Remote Control Commands
R&S
®
SMBV100A
774Operating Manual 1176.8016.02 ─ 17
Example:
LIST:DEXC:MODE IMP
selects that ASCII files with frequency and level value pairs are
imported and transferred into list mode lists.
LIST:DEXC:AFIL:SEL ‘/var/user/list.csv'
selects that ASCII file list.csv is imported.
LIST:DEXC:SEL ‘/var/user/list_imp'
selects that the ASCII file list.csv is imported into list mode
list list_imp.
Manual operation: See " Select Destination / Source - List Mode " on page 295
[:SOURce<hw>]:LIST:DWELl
<Dwell>
Sets the dwell time. The R&S SMBV generates the signal with the frequency / power
value pairs of each list entry for that particular period.
Parameters:
<Dwell> float
Range: 7E-4 to 100
Increment: 1E-4
*RST: 15E-3
Example:
LIST:DWEL 15
each setting in the list is retained for 15 ms.
Manual operation: See "Dwell Time - List Mode" on page 291
[:SOURce<hw>]:LIST:FREE?
Queries on the free storage space for list mode lists.
Return values:
<Free> integer
Range: 0 to INT_MAX
*RST: 0
Example:
LIST:FREE?
Usage: Query only
Response: 2147483647;1
[:SOURce<hw>]:LIST:FREQuency <Frequency>
Fills the FREQuency column of the selected list with data.
*RST does not affect data lists.