Remote Control Commands
R&S
®
SMBV100A
841Operating Manual 1176.8016.02 ─ 17
Return values:
<Result> 0 | 1 | RUNning | STOPped
0
Success
1
Fail
*RST: STOPped
Example:
TEST:ALL:STAR
Starts the self-test
TEST:ALL:RES?
Usage: Query only
Manual operation: See "Start Selftest" on page 133
:TEST:EIQMode
<EiqMode>
Triggers a connection test for testing the active external IQ devices.
Parameters:
<EiqMode> IQIN | IQOut
*RST: IQOut
:TEST<hw>:DIRect <HW_assembly>,<subadress>,<hex data string>
:TEST<hw>:DIRect? <HW_assembly>,<subadress>
The respective hardware assembly responds directly to the command; any safety
mechanisms are bypassed. This function is only available via remote control.
Example:
TEST:DIR 'SSYN',0,#H12345678
TEST:DIR? 'SSYN',0
Response: #H12345678
:TEST:BB:DATA:ERRor:RATE <ErrorRate>
Sets the error bit pattern of the BER/BLER test generator.
Parameters:
<ErrorRate> float
Range: 1E-6 to 1
Increment: 1E-6
*RST: 0.01
Manual operation: See "Error Rate" on page 205
:TEST:BB:DATA:FREQuency <Clock>
Sets the clock frequency for the BER/BLER test generator.
TEST Subsystem