Configuring the Internal Baseband Source
R&S
®
SMBV100B
174User Manual 1178.4460.02 ─ 03
They can be inserted after the TYPE tag in arbitrary order.
An example of the control list file contents could be:
{TYPE:SMU-CL}{COPYRIGHT:Rohde&Schwarz}
{DATE:2012-06-11;15:00:09}{HOP LIST:0:0;498:1;506:0}
{CW MODE LIST:0:0;380:1}{LEVATT LIST 3:0:0;464:1}
{BURST LIST:0:0;122:1;270:0;582:1;924:0}
{MARKER LIST 4:0:0;706:1;764:0}
{MARKER LIST 3:0:0;530:1;633:0}
{MARKER LIST 2:0:0;350:1;457:0}
{MARKER LIST 1:0:0;108:1;160:0}
{CONTROL LENGTH:1000}
The Figure 3-10 shows the representation of the created control list in the "Control
Data Editor".
Compare the displayed ramp values of "Marker 1" and the "Total List Length" with
the values in the corresponding tags.
Note: In the provided example, the tags have been separated by line breaks for
better reading.
See also Example "How to assign and activate control signals from a control list"
on page 175.
To create a control list in binary format
Generation of a control list in binary format is not necessary but possible.
Use the commands BB:DM:CLISt:... to generate a control list in binary format
(see "Handling List Files" on page 731 and the example in [:SOURce<hw>]:BB:
DM:CLISt:DATA on page 734).
See also Example "How to assign and activate control signals from a control list"
on page 175.
Using the Arbitrary Waveform Generator (ARB)