Configuring the RF Signal
R&S
®
SMBV100B
317User Manual 1178.4460.02 ─ 03
Table 6-3: Cross-reference between manual and remote control in Step / Extern Step modes (Sweep/
List)
Manual control mode:
"Step / Extern Step"
Remote commands
Rohde & Schwarz proprietary and [SCPI conform] value name
RF frequency sweep :TRIGger<hw>:FSWeep:SOURce SINGle [BUS] "Step" or
:TRIGger<hw>:FSWeep:SOURce EXTernal [EXTernal] for "Extern
Step"
:SOURce<hw>:SWEep:FREQuency:MODE STEP
RF level sweep :TRIGger<hw>:PSWeep:SOURce SINGle [BUS] "Step" or
:TRIGger<hw>:PSWeep:SOURce EXTernal [EXTenal] "Extern Step"
:SOURce<hw>:SWEep:POWer:MODE STEP
LF frequency sweep :TRIGger<hw>:LFFSweep:SOURce SINGle [BUS] "Step" or
:TRIGger<hw>:LFFSweep:SOURce EXTernal [EXTenal] "Extern Step"
:SOURce<hw>:LFOutput<ch>:SWEep:FREQuency:MODE STEP
List :SOURce<hw>:LIST:TRIGger:SOURce SINGle "Step" or
:SOURce<hw>:LIST:TRIGger:SOURce EXTernal "Extern Step"
:SOURce<hw>:LIST:MODE STEP
Extern Start/Stop mode (sweep)
Figure 6-11: Extern Start/Stop mode (sweep)
The instrument generates the signal continuously.
Trigger mode "Auto" and "Sweep > State = On" are prerequisite.
State "On" sets the signal to the start value, that is one of the following:
Sweep start frequency
Sweep start power
Frequency-power value pair of the selected index in the list
Starts signal generation with a trigger event.
Varying the RF Signal in List or Sweep Mode