Configuring the RF Signal
R&S
®
SMBV100B
318User Manual 1178.4460.02 ─ 03
Switches automatically to the next sweep step when the Dwell time has elapsed.
If the end value is reached, signal generation continues with the next sweep cycle.
Stops signal generation with the next external trigger event.
Starts the signal generation again with the next trigger event, beginning at the start
value.
"State = Off" stops the signal generation in sweep or list mode.
Trigger source: An externally applied trigger signal.
Table 6-4: Cross-reference between manual and remote control in Extern Start/Stop modes (sweep)
Manual control mode:
"Extern Start/Stop"
Remote commands
Rohde & Schwarz proprietary and [SCPI conform] value name
RF frequency sweep
:TRIGger<hw>:FSWeep:SOURce EAUTo [-]
:SOURce<hw>:SWEep:FREQuency:MODE AUTO
RF level sweep
:TRIGger<hw>:PSWeep:SOURce EAUTo [-]
:SOURce<hw>:SWEep:POWer:MODE AUTO
LF frequency sweep
:TRIGger<hw>:LFFSweep:SOURce EAUTo [-]
:SOURce<hw>:LFOutput<ch>:SWEep:FREQuency:MODE AUTO
Manual mode (Sweep/List)
The manual mode only applies to remote control. It is not visible in the graphical user
interface of the instrument and is described here for completeness.
Figure 6-12: Manual mode in remote control (sweep / list )
The instrument generates the signal in steps.
Starts signal generation with "State = On".
Switches to the next step with a user defined setting via remote control.
Varying the RF Signal in List or Sweep Mode