Configuring the RF Signal
R&S
®
SMBV100B
333User Manual 1178.4460.02 ─ 03
"Logarithmic"
Takes the value entered as a logarithmic value, i.e. as a constant
fraction of the current frequency in %.
Remote command:
[:SOURce<hw>]:SWEep[:FREQuency]:SPACing on page 886
[:SOURce<hw>]:LFOutput:SWEep[:FREQuency]:SPACing on page 849
Step Linear/Step Logarithmic
In "RF/LF Frequency Sweep" mode, sets the step width for the individual frequency
sweep steps. The value is added at each sweep step to the current frequency.
Depending on the current Spacing, you can enter either an absolute or logarithmic step
width.
See Chapter 6.10.2.1, "Correlating Parameters in Sweep Mode", on page 320.
"Step Linear"
The step width is a constant value in Hz.
"Step Logarithmic"
The step width is determined logarithmically in %, i.e. as a constant
fraction of the current frequency.
Remote command:
[:SOURce<hw>]:SWEep[:FREQuency]:STEP[:LINear] on page 888
[:SOURce<hw>]:SWEep[:FREQuency]:STEP:LOGarithmic on page 888
[:SOURce<hw>]:LFOutput:SWEep[:FREQuency]:STEP[:LINear] on page 850
[:SOURce<hw>]:LFOutput:SWEep[:FREQuency]:STEP:LOGarithmic
on page 849
6.10.5.3 Level Range Settings
Access:
1. Select "RF" > "Sweep/List" > "RF Level Sweep"
2. Select "Level Range".
Varying the RF Signal in List or Sweep Mode