Configuring the RF Signal
R&S
®
SMBV100B
385User Manual 1178.4460.02 ─ 03
State
Activates user correction.
Remote command:
[:SOURce<hw>]:CORRection[:STATe] on page 815
User Correction
Indicates the corrected level value for a specific frequency point.
Remote command:
[:SOURce<hw>]:CORRection:VALue? on page 814
UCOR Data
Accesses the standard "File Select" function of the instrument. The provided navigation
possibilities in the dialog are self-explanatory.
Files with user correction values are files with predefined file extension *.uco. When a
file is selected, the dialog indicates the filename.
You can create the file internally in the table editor or externally.
To select an existing file, select "Select List > navigate to the file *.uco > Select"
Use the general editor function to create internally new file or to edit an existing
one.
Use the standard file manager function to load externally created files to the instru-
ment.
Remote command:
[:SOURce]:CORRection:CSET:CATalog? on page 815
[:SOURce<hw>]:CORRection:CSET[:SELect] on page 814
[:SOURce]:CORRection:CSET:DELete on page 816
Edit UCOR Data
Opens the build-in table editor to define a new correction table or edit an existing one.
See also:
Chapter 6.10.7, "List Editor", on page 341
" Fill... " on page 344
Chapter 6.12.3.3, "Fill with Sensor", on page 389
6.12.3.2 List Editor
The "User Correction" and "List Mode" dialogs provide a build-in list editor for defining
the corresponding value pairs.
The table and navigation functions of the UCOR and list mode editors have the same
structure and are therefore summarized below. The access is specified for both modes,
the fields and functions are explained using the example of the list mode.