Configuring the RF Signal
R&S
®
SMBV100B
404User Manual 1178.4460.02 ─ 03
7. Close the dialog.
To create user correction data with an R&S NRP and the R&S SMBV100B
We assume, that the power sensor is connected, assigned and ready for operation.
1. Select "RF" > "RF Level" > "User Correction".
2. Select "UCOR Data > UCOR Data...".
3. Create a file: "New > Filename" and confirm with "OK".
4. Select the new file with "Select".
5. Select "Edit UCOR Data...".
6. Select "Edit > Fill".
The "Fill Table" dialog enables you to fill in the values of the columns automatically.
7. To fill in the frequency column:
a) Select "Select Column To Fill > Frequency / Hz".
b) Select "Range > e.g. 15" to determine the number of values.
c) Select "Start Value > e.g. 1.56 GHz".
d) Select "Increment Value > e.g. 1 MHz", to determine the frequency steps.
e) Select "Fill", to insert the frequency values.