Multi-Instrument Setups
R&S
®
SMBV100B
430User Manual 1178.4460.02 ─ 03
Multi Instrument Trigger
Sets the mode in that the instrument works.
To activate the mode, set "Multi Instrument > State = On".
"Master"
Provides its trigger signal to the slaves
"Slave"
Synchronizes to the supplied trigger signal from the master.
Observe the "Synchronisation State" indication for information if the
synchronization signal is found and synchronization is achieved.
Remote command:
:SCONfiguration:MULTiinstrument:MODE on page 664
Synchronisation State
If "Multi Instrument Trigger = Slave", indicates if synchronization is achieved.
"Sync"
Instrument is in sync.
It receives and decodes the synchronization and the trigger signal
from the master instrument.
"No Sync"
Synchronization is not achieved, because:
The baseband generators in the slaves or in the master instru-
ments are not activated, see "To start signal generation"
on page 433.
There is no physical connection (the synchronization signal is not
connected as required).
For details, see Chapter 8.3.1.2, "How to Connect and Configure
Instruments in Master-Slave Mode", on page 430.
The slave instrument does not detect the synchronization signal.
Remote command:
:SCONfiguration:MULTiinstrument:TRIGger:SYNChronization?
on page 665
Global Connector Settings
Provides a quick access to the related global connector settings.
With "State = On", the synchronization signal is assigned per default as follows:
"Multi Instrument Trigger > Master"
"Connector = User 5": "Direction = Output", "Signal = Baseband Sync Out"
"Multi Instrument Trigger > Slave"
"Connector = User 5": "Direction = Output", "Signal = Baseband Sync Out"
"Connector = User 4": "Direction = Input", "Signal = Baseband Sync In"
See also:
Chapter 8.3.1.2, "How to Connect and Configure Instruments in Master-Slave
Mode", on page 430.
See Chapter 10.2, "Configuring the Global Connectors", on page 515.
8.3.1.2 How to Connect and Configure Instruments in Master-Slave Mode
Proceed as follows:
Connect the instruments and distribute the synchronization signal and the refer-
ence frequency
Generating Time-Aligned Baseband Signals