Multi-Instrument Setups
R&S
®
SMBV100B
461User Manual 1178.4460.02 ─ 03
Sample Rate ← Channel Table
Indicates the used sample rate per channel.
For information on the max. sample rate and the aggregated sample rate, depending
on the number of active channels, observe the indication Aggregated Link Sample
Rate.
Remote command:
[:SOURce<hw>]:BBIN:CHANnel<ch0>:SRATe on page 685
Crest Factor ← Channel Table
If DIG IQ Auto Setting > "Off", sets the crest factor per channel.
Remote command:
[:SOURce<hw>]:BBIN:CHANnel<ch0>:POWer:CFACtor on page 684
Peak Level ← Channel Table
If DIG IQ Auto Setting > "Off", sets the peak level per channel. The value is set relative
to full scale of 0.5 V (in terms of dB full scale).
Remote command:
[:SOURce<hw>]:BBIN:CHANnel<ch0>:POWer:PEAK on page 684
Level ← Channel Table
Indicates the estimated RMS level acquired with "Auto Level Set".
Remote command:
[:SOURce<hw>]:BBIN:CHANnel<ch0>:POWer:RMS on page 685
BB ← Channel Table
Activates the channel. Only one channel can be activated at a time. Toggle the state of
the channel to test all of them subsequently.
Remote command:
[:SOURce<hw>]:BBIN:CHANnel<ch0>:BB:STATe on page 684
Aggregated Link Sample Rate
Indicates the aggregated sample rate and the maximum sample rate, where:
The aggregated sample rate is the sum of the sample rates of all active channels.
It can not exceed the maximum sample rate.
The R&S SMBV100B supports two standard max. sample rate values.
This parameter indicates the currently selected one, where the max. sample rate is
selected depending on the capabilities of the transmitter.
For more information, see data sheet.
Remote command:
[:SOURce<hw>]:BBIN:SRATe:SUM? on page 685
[:SOURce<hw>]:BBIN:SRATe:MAX? on page 685
DIG IQ Auto Setting
Activates an automatic adjustment of the baseband input signal.
The R&S SMBV100B receives PEP, level and crest factor values directly from the con-
nected transmitter and recognizes changes automatically.
Digital Baseband Input Settings