Remote Control Commands
R&S
®
SMBV100B
631User Manual 1178.4460.02 ─ 03
Example:
CALibration:ALL:TEMP?
// "+39.00 K"
Usage: Query only
Manual operation: See "Temperature Offset Since Last Full Adjustment"
on page 935
:CALibration<hw>:ALL:TIME?
Queries the time elapsed since the last full adjustment.
Return values:
<Time> string
Example:
CAL:ALL:TIME?
// "22 days"
Usage: Query only
Manual operation: See "Time Since Last Full Adjustment" on page 935
:CALibration:DATA:FACTory:DATE?
Queries the date of the last factory calibration.
Return values:
<Date> string
Example:
CAL:DATA:FACT:DATE?
// "2016-01-01"
Usage: Query only
Manual operation: See "Last Factory Calibration" on page 942
:CALibration<hw>:CONTinueonerror <State>
Continues the calibration even though an error was detected. By default adjustments
are aborted on error.
Suffix:
<hw>
.
[1]
Optional suffix
Parameters:
<State> 0 | 1 | OFF | ON
*RST: n.a. (factory preset: 0)
Example:
CAL:CONT ON
// Continues calibration after an error
CALibration Subsystem