Remote Control Commands
R&S
®
SMBV100B
647User Manual 1178.4460.02 ─ 03
Suffix Value range Description
OUTPut<hw>
[1] Optional suffix
:OUTPut:ALL[:STATe]..................................................................................................... 647
:OUTPut<hw>[:STATe]....................................................................................................647
:OUTPut<hw>[:STATe]:PON............................................................................................647
:OUTPut<hw>:AMODe................................................................................................... 648
:OUTPut<hw>:IMPedance?.............................................................................................648
:OUTPut<hw>:AFIXed:RANGe:LOWer?...........................................................................648
:OUTPut<hw>:AFIXed:RANGe:UPPer?............................................................................648
:OUTPut<hw>:PROTection:CLEar................................................................................... 649
:OUTPut<hw>:PROTection:TRIPped?.............................................................................. 649
:OUTPut:ALL[:STATe] <State>
Activates the RF output signal of the instrument.
Parameters:
<State> 0 | 1 | OFF | ON
*RST: n.a. (factory preset: 0)
Example:
OUTPut:ALL:STATe 0
:OUTPut<hw>[:STATe] <State>
Activates the RF output signal.
Parameters:
<State> 0 | 1 | OFF | ON
*RST: 0
Example:
OUTP ON
Activates the RF output.
Manual operation: See "RF State/RF ON" on page 288
:OUTPut<hw>[:STATe]:PON <Pon>
Defines the state of the RF output signal when the instrument is switched on.
Parameters:
<Pon> OFF | UNCHanged
*RST: n.a. (factory preset: UNCHanged)
Example:
OUTP:PON OFF
The RF output is deactivated when the instrument is switched
on.
Manual operation: See " Power-On State " on page 513
OUTPut Subsystem