Remote Control Commands
R&S
®
SMBV100B
672User Manual 1178.4460.02 ─ 03
:SCONfiguration:EXTernal:REMote:SCAN
Scans the network for connected instruments.
Example: See Example "Connecting and configuring external instrument"
on page 662.
Usage: Event
Manual operation: See "Detect/Scan" on page 450
:SCONfiguration:EXTernal:REMote:SCAN:STATe?
Queries if scanning is performed or not.
To start the scanning process, use the command :SCONfiguration:EXTernal:
REMote:SCAN on page 672.
Return values:
<ScanState> 0 | 1 | OFF | ON
1
Scanning process running
0
Not scanning
Example: See Example "Connecting and configuring external instrument"
on page 662.
Usage: Query only
Manual operation: See "Detect/Scan" on page 450
:SCONfiguration:EXTernal:REMote:LIST?
Lists all available instruments, found by the :SCONfiguration:EXTernal:REMote:
SCAN command.
Return values:
<InstrNames> String
String with symbolic names and/or alias names
Example: See Example "Connecting and configuring external instrument"
on page 662.
Usage: Query only
Manual operation: See "External Instrument" on page 451
:SCONfiguration:EXTernal:REMote:ADD <InstrName>, <HwChan>,
<TcpIporUsbAddr>[, <RfPathNumber>]
Adds manually an external instrument to the list of available instruments.