Remote Control Commands
R&S
®
SMBV100B
705User Manual 1178.4460.02 ─ 03
SOURce:PULM:STATe 1
OUTPut1:STATe 1
The following commands are available:
[:SOURce<hw>]:PULM:MODE........................................................................................ 705
[:SOURce<hw>]:PULM:TRIGger:MODE...........................................................................705
[:SOURce]:PULM[:INTernal][:TRAin]:TRIGger:IMMediate................................................... 706
[:SOURce<hw>]:PULM:PERiod....................................................................................... 706
[:SOURce<hw>]:PULM:WIDTh........................................................................................706
[:SOURce<hw>]:PULM:DELay........................................................................................ 706
[:SOURce<hw>]:PULM:DOUBle:DELay............................................................................707
[:SOURce<hw>]:PULM:DOUBle:WIDTh........................................................................... 707
[:SOURce<hw>]:PULM:DOUBle:STATe............................................................................ 707
[:SOURce<hw>]:PULM:STATe.........................................................................................707
[:SOURce<hw>]:PULM:SOURce..................................................................................... 707
[:SOURce<hw>]:PULM:TTYPe........................................................................................708
[:SOURce<hw>]:PULM:OUTPut:VIDeo:POLarity...............................................................708
[:SOURce<hw>]:PULM:POLarity..................................................................................... 708
[:SOURce<hw>]:PULM:IMPedance..................................................................................709
[:SOURce<hw>]:PULM:MODE <Mode>
Selects the mode for the pulse modulation.
Parameters:
<Mode> SINGle | DOUBle
SINGle
Generates a single pulse.
DOUBle
Generates two pulses within one pulse period.
*RST: SINGle
Example: See Example "Perform pulse modulation" on page 704.
Manual operation: See "Pulse Mode" on page 358
[:SOURce<hw>]:PULM:TRIGger:MODE <Mode>
Selects a trigger mode - auto, single, external, external single or external gated - for
generating the modulation signal.
Parameters:
<Mode> AUTO | EXTernal | EGATe | SINGle | ESINgle
*RST: AUTO
Example: See Example "Perform pulse modulation" on page 704.
Manual operation: See "Trigger Mode" on page 359