Remote Control Commands
R&S
®
SMBV100B
823User Manual 1178.4460.02 ─ 03
Usage: Query only
Manual operation: See "Mandatory LO IN Frequency" on page 308
[:SOURce<hw>]:FREQuency:LOSCillator:MODE <Mode>
Selects the mode of the local oscillator coupling.
Parameters:
<Mode> INTernal | EXTernal
*RST: INTernal
Example: See Example "Set up of LO coupling and LO level adjustment
with three Rohde & Schwarz signal generators" on page 821.
Manual operation: See "Mode" on page 307
[:SOURce<hw>]:FREQuency:LOSCillator:OUTPut:FREQuency?
Queries the current frequency of the local oscillator at the [LO Out] connector.
Return values:
<Frequency> float
Range: 100E3 to 20E9
Increment: 0.01
*RST: 0
Example: See Example "Set up of LO coupling and LO level adjustment
with three Rohde & Schwarz signal generators" on page 821.
Usage: Query only
Manual operation: See "LO OUT Frequency" on page 308
[:SOURce<hw>]:FREQuency:LOSCillator:OUTPut:STATe <State>
Activates the LO output.
Parameters:
<State> 0 | 1 | OFF | ON
*RST: 0
Example: See Example "Set up of LO coupling and LO level adjustment
with three Rohde & Schwarz signal generators" on page 821.
Manual operation: See "Out State " on page 307