Remote Control Commands
R&S
®
SMBV100B
875User Manual 1178.4460.02 ─ 03
[:SOURce<hw>]:POWer[:LEVel][:IMMediate]:RCL <Rcl>
Determines whether the current level is retained or if the stored level setting is adopted
when an instrument configuration is loaded.
Parameters:
<Rcl> INCLude | EXCLude
INCLude
Takes the current level when an instrument configuration is loa-
ded.
EXCLude
Retains the current level when an instrument configuration is
loaded.
*RST: INCLude
Example:
POW:RCL INCL
Takes the level value from an instrument configuration loaded
with command *RCL.
Manual operation: See "Exclude Level" on page 484
[:SOURce<hw>]:POWer[:LEVel][:IMMediate][:AMPLitude] <Amplitude>
Sets the RF level applied to the DUT.
To activate the RF output use command :OUTPut<hw>[:STATe] ("RF On"/"RF Off").
The following applies POWer = RF output level + OFFSet, where:
●
POWer is the values set with [:SOURce<hw>]:POWer[:LEVel][:
IMMediate][:AMPLitude]
●
RF output level is set with [:SOURce<hw>]:POWer:POWer
●
OFFSet is set with [:SOURce<hw>]:POWer[:LEVel][:IMMediate]:OFFSet
Parameters:
<Amplitude> float
The following settings influence the value range:
OFFSet set with the command [:SOURce<hw>]:POWer[:
LEVel][:IMMediate]:OFFSet
Numerical value
Sets the level
UP|DOWN
Varies the level step by step.
The level is increased or decreased by the value set with the
command [:SOURce<hw>]:POWer:STEP[:INCRement].
Range: (Level_min + OFFSet) to (Level_max + OFFStet)
*RST: -30
Default unit: dBm
Example:
POWer -30
Sets the RF level
SOURce Subsystem