Remote Control Commands
R&S
®
SMBV100B
884User Manual 1178.4460.02 ─ 03
[:SOURce<hw>]:SWEep:POWer:DWELl <Dwell>
Sets the dwell time for a level sweep step.
Parameters:
<Dwell> float
Range: 0.001 to 100
Increment: 100E-6
*RST: 0.01
Manual operation: See " Dwell Time " on page 330
[:SOURce<hw>]:SWEep:POWer:MODE
<Mode>
Sets the cycle mode for the level sweep.
Parameters:
<Mode> AUTO | MANual | STEP
AUTO
Each trigger triggers exactly one complete sweep.
MANual
The trigger system is not active. You can trigger every step indi-
vidually with the command [:SOURce<hw>]:POWer:MANual.
The level value increases at each step by the value that you
define with [:SOURce<hw>]:POWer:STEP[:INCRement].
Values directly entered with the command [:SOURce<hw>]:
POWer:MANual are not taken into account.
STEP
Each trigger triggers one sweep step only. The level increases
by the value entered with [:SOURce<hw>]:POWer:STEP[:
INCRement].
*RST:
AUTO
Example: See Example "Setup an RF frequency or power sweep"
on page 882.
Manual operation: See "Mode" on page 328
[:SOURce<hw>]:SWEep:POWer:POINts <Points>
Sets the number of steps within the RF level sweep range.
See Chapter 6.10.2.1, "Correlating Parameters in Sweep Mode", on page 320.
Parameters:
<Points> integer
Range: 2 to Max
Example: See Example "Setup an RF frequency or power sweep"
on page 882.
SOURce Subsystem