Remote Control Commands
R&S
®
SMBV100B
885User Manual 1178.4460.02 ─ 03
[:SOURce<hw>]:SWEep:POWer:SPACing:MODE?
Queries the level sweep spacing. The sweep spacing for level sweeps is always linear.
Return values:
<Mode> LINear
*RST: LINear
Example:
SWE:POW:SPAC:MODE?
queries the sweep spacing for a level sweep at RF output.
Result: "LIN"
linear spacing
Usage: Query only
[:SOURce<hw>]:SWEep:POWer:STEP[:LOGarithmic] <Logarithmic>
Sets a logarithmically determined step size for the RF level sweep. The level is
increased by a logarithmically calculated fraction of the current level.
See Chapter 6.10.2.1, "Correlating Parameters in Sweep Mode", on page 320.
Parameters:
<Logarithmic> float
The unit dB is mandatory.
Range: 0.01 to 139 dB
Increment: 0.01
*RST: 1
Default unit: dB
Example: See Example "Setup an RF frequency or power sweep"
on page 882.
Manual operation: See " Step " on page 334
[:SOURce<hw>]:SWEep[:FREQuency]:DWELl <Dwell>
Sets the dwell time for a frequency sweep step.
Parameters:
<Dwell> float
Range: 0.001 to 100
Increment: 100E-6
*RST: 0.01
Manual operation: See " Dwell Time " on page 330
SOURce Subsystem