EasyManua.ls Logo

Rohde & Schwarz SMBV100B - Page 896

Rohde & Schwarz SMBV100B
1030 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Control Commands
R&S
®
SMBV100B
894User Manual 1178.4460.02 ─ 03
:SYSTem:ERRor[:NEXT]?
Queries the error/event queue for the oldest item and removes it from the queue.
Return values:
<Next> string
Error/event_number,"Error/event_description>[;Device-depend-
ent info]"
Error number and a short description of the error.
If the queue is empty, the response is 0,"No error"
Positive error numbers are instrument-dependent. Negative
error numbers are reserved by the SCPI standard.
Volatile errors are reported once, at the time they appear. Identi-
cal errors are reported repeatedly only if the original error has
already been retrieved from (and hence not any more present in)
the error queue.
Example:
SYST:ERR?
Queries the oldest entry in the error queue.
Response: 0, 'no error'
No errors have occurred since the error queue was last read out.
Usage: Query only
Manual operation: See "Static Errors/Error History" on page 950
:SYSTem:ERRor:GNEXt?
Similar to :SYSTem:ERRor[:NEXT]?, but queries the next entry from the global per-
sistant error/event queue.
Return values:
<NextGlobalError> string
Error/event number, "Error/event description> [;Device depend-
ent info]"
An error number and a short description of the error.
Positive error numbers are instrument-dependent. Negative
error numbers are reserved by the SCPI standard.
Example:
SYST:ERR:GNEX?
Returns the next error message from the global error queue.
Usage: Query only
:SYSTem:ERRor:HISTory:CLEar
Clears the error history.
Example:
See Example "Querying the error queue" on page 890
Usage: Event
Manual operation: See "Clear History" on page 950
SYSTem Subsystem

Table of Contents

Related product manuals