Remote Control Commands
R&S
®
SMBV100B
909User Manual 1178.4460.02 ─ 03
Usage: Query only
Manual operation: See "Versions" on page 944
:SYSTem:SRData?
Queris the SCPI recording data from the internal file.
This feature enables you to transfer an instrument configuration to other test environ-
ments, as e.g. laboratory virtual instruments.
Return values:
<FileData> block data
Example:
SYSTem:SRData?
// #3118:SOURce1:ROSCillator:SOURce EXT
// :SOURce1:FREQuency:CW 4000000000
// :SOURce1:FREQuency:OFFSet 1000000
// :SOURce1:AM1:STATe 1
// :OUTPut1:STATe 1
Usage: Query only
:SYSTem:STARtup:COMPlete?
Queries if the startup of the instrument is completed.
Return values:
<Complete> 0 | 1 | OFF | ON
*RST: 0
Example:
SYST:STAR:COMP?
Response: 1
the startup of the instrument is completed.
Usage: Query only
:SYSTem:DATE <Year>, <Month>, <Day>
Queries or sets the date for the instrument-internal calendar.
This is a password-protected function. Unlock the protection level 1 to access it, see :
SYSTem:PROTect<ch>[:STATe].
Parameters:
<Year> integer
<Month> integer
Range: 1 to 12
<Day> integer
Range: 1 to 31
SYSTem Subsystem