Remote Control Commands
R&S
®
SMBV100B
918User Manual 1178.4460.02 ─ 03
:TEST:BB:GENerator:ARBitrary....................................................................................... 920
:TEST:BB:GENerator:FREQuency................................................................................... 920
:TEST:BB:GENerator:GAIN.............................................................................................920
:TEST:BB:GENerator:SOURce........................................................................................ 921
:TEST:BB:GENerator:STATe............................................................................................921
:TEST<hw>:ALL:STARt.................................................................................................. 921
:TEST<hw>:ALL:RESult?................................................................................................921
:TEST:BASeband?
Queries the result of the baseband selftest.
Return values:
<TestBbError> 0 | 1 | RUNning | STOPped
Example:
:TEST:BASeband?
// 0
// the test succeeded
Usage: Query only
Manual operation: See "Execute Baseband Selftest" on page 937
:TEST:BB:BNC:CONNection?
Queries the BNC connection test result.
This is a password-protected function. Unlock the protection level 1 to access it, see :
SYSTem:PROTect<ch>[:STATe].
Return values:
<TestStatus> 0 | 1 | RUNning | STOPped
Example:
SYSTem:PROTect1 0,123456
TEST:BB:BNC:CONNection?
// RUN
// test is running
Usage: Query only
Manual operation: See "Execute BNC Connection Test" on page 936
:TEST:BB:BNC:SOURce <BncSource>
Selects the BNC connection test source.
This is a password-protected function. Unlock the protection level 1 to access it, see :
SYSTem:PROTect<ch>[:STATe].
Parameters:
<BncSource> AUTO | USER1 | USER2 | USER3 | USER4 | USER5
*RST: USER1