R&S SMC100A User Correction
1411.4060.32 4.65 E-4
Filling the Correction List with Power Sensor Measurement Data
The level correction values for the user correction list can be acquired by means of R&S NRP power
sensors. The NRP sensor isconnected to one of the USB interfaces. Configuration of the connection is
performed in the Power Sensor menu (see "Power Viewer"). The filling of the ucor list with
measurement data is performed in the ucor list editor (see above "Edit/New User Correction Data - User
Correction").
In the editor, the frequencies for which the correction values are to be acquired are entered in the
frequency column (either manually or by means of the Fill… menu).
Note:
Do not save the list at this point, because the frequency entries are lost as long as there are no
entries for the level column also. In the following these entries are automatically acquired by the
connected power sensor.
All level correction values for the given frequency values are measured using the power sensor and
automatically filled in the selected list. The list is automatically stored and recalled again after filling.
The Fill with Sensor button of the Edit User Correction Data menu opens the associated menu.
The menu indicates the relevant generator settings. The Execute button is only enabled if a sensor is
detected and the user correction list contains at least one frequency value.
Remote-control command:
SOUR:CORR:CSET:DATA:SENS:POW:SONC