Command Description R&S SMC100A
1411.4060.32 6.28 E-1
SENSe[:POWer]:Offset -100 … 100
The command enters a level offset which is added to the measured level value after activation with
command SENSe[:POWer]:Offset:STATe ON. This allows e.g. an attenuator in the signal path
to be taken into account.
Example:
SENS:POW:OFFS 10.0
'sets a level offset of 10 dB
*RST value Resolution SCPI
0.0 dB - -
SENSe[:POWer]:Offset:STATe ON | OFF
The command activates the addition of the level offset to the measured value. The level offset value
is set with command SENSe[:POWer]:Offset.
Example:
SENS:POW:OFFS 0.4dB
'sets a level offset of 0.4 dB
SENS:POW:OFFS:STAT ON
'a level offset of 0.4 dB is added to the measured value.
*RST value Resolution SCPI
OFF - -
SENSe[:POWer]:SNUMber?
The command queries the serial number of the sensor.
The command is a query command and therefore has no *RST value.
Example:
"SENS:SNUM?"
'queries the serial number.
*RST value Resolution SCPI
- - Device specific
SENSe[:POWer]:SOURce A | USER
The command selects the signal source for the measurement. The RF signal (A) or a user defined
source can be selected. The frequency of the user-defined source has to be entered with command
SENSe:POWer:FREQuency.
Example:
SENS:SOUR A
'selects the RF signal as measurement source. The RF frequency is used as
the measurement frequency of the sensor and the corresponding correction
factor is used. The level setting of the instrument serves as reference level of
the measurement.
*RST value Resolution SCPI
A - Device specific