R&S SMC100A Command Description
1411.4060.32 6.41 E-1
[SOURce:]CORRection[:STATe] ON | OFF
The command activates/deactivates level correction. Level correction is performed using the table
which has been selected with the command CORRection:CSET:SELect.
Example:
SOUR:CORR:CSET 'ucor1
'selects the table ucor1.
SOUR:CORR ON
'activates user correction.
*RST value Resolution Options SCPI
OFF - Device-specific
[SOURce:]CORRection:VALue?
The command requests the current value for user correction.
This command is a query command and has no *RST value.
Example:
CORR:VAL?
'queries the value currently used for level correction.
Response: -3
'the correction value is - 3dB.
*RST value Resolution Options SCPI
- - Device-specific