Command Description R&S SMC100A
1411.4060.32 6.50 E-1
The keyword SOURce is optional and can be omitted.
Command Parameters Default
unit
Remark
[SOURce:]LFOutput:FREQuency Min max Hz
See data
sheet
[SOURce:]LFOutput:FREQuency:MANual 0.1 Hz ... 1 MHz Hz
[SOURce:]LFOutput:FREQuency:MODE CW | FIXed | SWEep
[SOURce:]LFOutput:FREQuency:STARt 0.1 Hz ... 1 MHz Hz
[SOURce:]LFOutput:FREQuency:STOP 0.1 Hz ... 1 MHz Hz
[SOURce:]LFOutput[:STATe] ON | OFF
[SOURce:]LFOutput:SWEep[:FREQuency]:DWELl 2 ms ... 10 s s
[SOURce:]LFOutput:SWEep[:FREQuency]:EXECute -
[SOURce:]LFOutput:SWEep[:FREQuency]:MODE AUTO | MANual | STEP
[SOURce:]LFOutput:SWEep[:FREQuency]:POINts <numeric_value>
[SOURce:]LFOutput:SWEep[:FREQuency]:SHAPe SAWTooth | TRIangle
[SOURce:]LFOutput:SWEep[:FREQuency]:SOURce LF1 | LF2
[SOURce:]LFOutput:SWEep[:FREQuency]:SPACing LINear | LOGarithmic
[SOURce:]LFOutput:SWEep[:FREQuency]:STEP[:LINear] 0 ... (STOP - STARt) Hz
[SOURce:]LFOutput:SWEep[:FREQuency]:STEP:LOGarithmic 0.01 ... 50 PCT PCT
[SOURce:]LFOutput:VOLTage 0 V ... 4 V V
SOURce-LFOutput - LF-Frequency[SOURce:]LFOutput:FREQuency 0.001 V... 4 V
The command sets the frequency of the LF signal for CW mode (:SOUR:MODE CW mode). The
setting is valid for all analog modulations (AM/FM/PhiM) with internal modulation source and for the
LF Output.
In Sweep mode (SOUR:LFO:FREQ:MODE SWE), the frequency is linked to the sweep frequency. "
Example:
LFO2:FREQ 5kHz
'sets the frequency of the LF generator 2 signal to 5 kHz'.
*RST value Resolution Options Correlation SCPI
1 kHz 0.1 Hz
LFO:FREQ for LFO:FREQ:MODE
SWE linked to sweep frequency
Compliant