EasyManua.ls Logo

Rohde & Schwarz SMC100A - KBOard Subsystem

Rohde & Schwarz SMC100A
352 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
R&S SMC100A Command Description
1411.4060.32 6.53 E-1
[SOURce:]LFOutput:SWEep[:FREQuency]:DWEL 2 ms...10 s
The command sets the dwell time for each frequency step of the sweep.
Note:
It is recommended to switch off the GUI update for optimum sweep performance especially with
short dwell times (SYSTem:DISPlay:UPDate OFF).
Example:
LFO:SWE:DWEL 20 ms
'sets a dwell time of 20 ms.
*RST value Resolution Options SCPI
10 ms 0.1 ms - Device-specific
[SOURce:]LFOutput:SWEep[:FREQuency]:EXECute
The command immediately starts an LF sweep. The command is only valid for sweep mode Single
(SOURce:LFOutput:SWEep:FREQuency:MODE SINGle). The command corresponds to the
manual-control command EXECUTE SINGLE SWEEP.
This command triggers an event and therefore has no *RST value.
Example:
LFO:SWE:MODE SING
'sets the single cycle mode of the LF sweep.
LFO:SWE:EXEC
'starts one cycle of the LF sweep.
*RST value Resolution Options SCPI
- - - Device-specific

Table of Contents

Other manuals for Rohde & Schwarz SMC100A

Related product manuals