EasyManua.ls Logo

Rohde & Schwarz SMC100A - Page 267

Rohde & Schwarz SMC100A
352 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
R&S SMC100A Command Description
1411.4060.32 6.65 E-1
[SOURce:]POWer:MODE CW|FIXed | SWEep
The command sets the instrument operating mode and therefore also the commands used to set the
output level.
Parameters: CW|FIXed
T
he instrument operates at a constant level. CW and FIXed are synonyms.
The output level is set with :SOURce:POWer.
SWEep
The instrument operates in SWEep mode. The level is set using the
commands SOURce:POWer:STARt; STOP; MANual.
Example:
POW:MODE SWEep
'sets the SWEep mode. The settings under SOURce:POW:STARt; STOP;
MANual become effective.
*RST value Resolution Options Correlation SCPI
CW - Compliant
[SOURce:]POWer:STARt Minimum level ... Maximum level
The command sets the RF start level in Sweep mode.
As with the Level value entered in the RF Level menu, the Offset value is also taken into
consideration with this command.
The specified value range is therefore only effective if :SOURce:POWer:Offset is set to 0. The
value range for other Offset values can be calculated using the following formula:
Minimum level + Offset ...Maximum level + Offset
Parameters: Minimum level ... Maximum level
The value range for the level setting varies according to the instrument model.
The values are given in the data sheet.
Example:
POW:STAR -20 dBm
'sets the start level for the level sweep to -15 dBm.
*RST value Resolution Options SCPI
-30 dBm 0.01 dB Compliant

Table of Contents

Other manuals for Rohde & Schwarz SMC100A

Related product manuals