R&S SMC100A Command Description
1411.4060.32 6.69 E-1
[SOURce:]PULM:MODE DOUBle | SINGle
The command selects the mode for pulse generation.
Parameters: SINGle
A single pulse is generated in one pulse period.
DOUBle
Two pulses are generated in one pulse period.
Example:
PULM:MODE DOUB
'double-pulse mode is enabled.
*RST value Resolution Options SCPI
SINGle Device-specific
[SOURce:]PULM:PERiod 5Ts .. 85.s | 100ns .. 85 s
The command sets the period of the generated pulse. The period determines the repetition
frequency of the internal signal.
Example:
PULM:PER 220 us
'the pulse period is 220 us.
*RST value Resolution Options SCPI
1 ms 1us | 20 ns Device-specific
[SOURce:]PULM:POLarity NORMal | INVerted
The command sets the polarity between modulating and modulated signal. This command is only
effective for an external modulation signal.
Parameters: NORMal
The RF signal is suppressed during the pulse pause.
INVerted
The RF signal is suppressed during the pulse.
Example:
PULM:SOUR EXT
'selects the external modulation source.
PULM:POL INV
'selects inverted polarity.
*RST value Resolution Options SCPI
NORMal - Device-specific