Command Description R&S SMC100A
1411.4060.32 6.80 E-1
[SOURce:]SWEep:POWer:EXECute
The command triggers a sweep.The command is only valid for sweep mode Single
(SOURce:SWEep:POWer:MODE SINGle). The command corresponds to the manual-control
command EXECUTE SINGLE SWEEP.
The command triggers an event and therefore has no query form and no *RST value.
Example:
SWE:POW:MODE SING
'sets the single cycle mode of the level sweep.
SWE:POW:EXEC
'triggers a level sweep.
*RST value Resolution Options SCPI
- - Device-specific
[SOURce:]SWEep:POWer:MODE AUTO | MANual | STEP
The command sets the cycle mode of the level sweep.
Parameters: AUTO
Each trigger triggers exactly one complete sweep.
MANual
The trigger system is not active. Each level step of the sweep is triggered
individually, either by varying the Current Level value using the rotary knob
under manual control or by means of a POW:MAN command under remote
control. With manual control, the level increases or decreases (depending on
the direction of the rotary encoder) by the value specified under
SOUR:SWE:POW:STEP. With remote control, the level increases by the value
specified under SWEep:POW:STEP which each sent :POW:MAN command,
irrespective the value entered there.
STEP
Each external trigger triggers one sweep step only (Mode Extern Step). The
level increases by the value entered under :SWEep:POWer:STEP.
Example:
SWE:POW:MODE AUTO
'selects Mode Auto for a level sweep.
*RST value Resolution Options SCPI
AUTO - Device-specific