Command Description R&S SMC100A
1411.4060.32 6.96 E-1
SYSTem:KLOCk ON | OFF
The command (K
eyboard LOCk) disables the front panel keyboard of the R&S Signal Generator
including the
LOCAL key, or enables it again (OFF).
Example:
SYST:KLOC ON
'activates the keyboard lock. The keyboard cannot be operated again until it
has been enabled with SYST:KLOC OFF.
*RST value Resolution Options SCPI
OFF - Compliant
SYSTem - Language
SYSTem:LANGuage?
The command queries the remote control command set.
The command is a query and therefore has no *RST value.
Example:
SYST:LANG?
Response "SCPI"
'the SCPI command set is used
*RST value Resolution Options SCPI
- - Device specific
SYSTem:PRESet
The command triggers an instrument reset. It has the same effect as the
PRESET key on the front
panel and the *RST command.
The command triggers an event and therefore has no *RST value and no query form.
Example:
SYST:PRES
'all instrument settings (also those that are not currently active) are reset to
their default values.
*RST value Resolution Options SCPI
- - Compliant