Alphabetical List of Device-Specific Error Messages R&S SMC100A
1411.4060.32 9.12 E-1
Alphabetical List of Device-Specific Error Messages
The following list contains all error messages specific of the instrument in alphabetical order. The
positive error codes mark the errors specific of the instrument.
The error text being entered into the error/event queue or being displayed is printed in bold face on the
left together with the error code. Below the error text, there is an explanation as to the respective error.
Note:
The index provides a list of the error messages sorted according to their error codes.
Extern reference out of range or disconnected (50)
External reference is selected but no external signal is applied or the signal is out of range.
Remedy:
Check the selected reference signal source (internal or external) in the
Reference Oscillator menu (Setup). Change setting to 'internal' if no approriate
external source is available.
SCPI:
Device-specific error - sets bit 3 in the ESR register
Adjustment data invalid (183)
Adjustment data are invalid and must be restored.
Remedy:
The adjustment data have to be generated again by an internal
or external adjustment or to be loaded into the instrument.
SCPI:
Device-specific error - sets bit 3 in the ESR register
Adjustment data missing (182)
Adjustment data are missing.
Remedy:
The adjustment data have to be generated first by an internal or
external adjustment or to be loaded into the instrument.
SCPI:
Device-specific error - sets bit 3 in the ESR register