Instrument Function
R&S
®
SMC100A
96Operating Manual 1411.4060.32 ─ 09
In addtion, you can also use a Bluetooth connection for remote control via the serial
interface. The settings are effective for both interfaces (see also Chapter 5.2.3.15,
"Security", on page 99 ).
Baud Rate ← RS232 using USB adapter
Sets the baudrate for the serial remote control interface.
Remote command:
:SYSTem:COMMunicate:SERial:BAUD on page 356
Parity ← RS232 using USB adapter
Sets the parity for the serial remote control interface.
Remote command:
:SYSTem:COMMunicate:SERial:PARity on page 356
Stop Bits ← RS232 using USB adapter
Sets the number of stop bits for the serial remote control interface.
Remote command:
:SYSTem:COMMunicate:SERial:SBITs on page 357
Visa Resource Strings
Displays the visa resource strings, used for remote control of the instrument. Each
interface requires an individual unique address, to identify the instrument for remote
control.
Remote command:
:SYSTem:COMMunicate:HISLip:RESource? on page 355
:SYSTem:COMMunicate:NETWork:RESource? on page 355
:SYSTem:COMMunicate:SOCKet:RESource? on page 357
:SYSTem:COMMunicate:GPIB:RESource? on page 355
:SYSTem:COMMunicate:USB:RESource? on page 355
:SYSTem:COMMunicate:SERial:RESource? on page 356
Goto Local
Switches the instrument to operate in local control mode.
Switching from remote to local control mode can be also done with one of the following
actions:
●
manually with the LOCAL key on the front panel
●
with the interface command >L via the remote control interface
●
with the key combination CTRL + Q.
Remote command:
>L
5.2.3.13 Instrument Emulations
It is also possible to remotely control the R&S SMC via the command set of another
signal generator, as for example of an HP generator. With this function you can, for
example, replace a signal generator with an R&S SMC in an automated test setup,
without adjusting the command scripts used.
General Instrument Settings